3-Pars Hematology Analyser Control

3-Pars Hematology Analyser Control

3- Pars Hematologia Analyser Imperium est tri-gradum referens materiam praeparatam ex suspensionibus sanguinis e cellulis stabiliendis aestimandis accurate et praecisionem calculis hematologiae automated.

Product Online

3- Pars Hematologia Analyser Imperium est tri-gradum referens materiam praeparatam ex suspensionibus sanguinis e cellulis stabiliendis aestimandis accurate et praecisionem calculis hematologiae automated.

 

Quid sunt Hematologia Imperium / Calibrators?

 

Haematologia controlla stabiliuntur suspensiones cellularum cum assignatis valoribus scoporum adhibitis ad monitorem exercitationis instrumenti perficiendi et praecisionis analyticae. Dissimiles calibratores, qui instrumenta uncinis et factores calibrationis baselines accommodant, moderantur ut exempla referentia independentes agunt ut errores systemicos egisse, errores fluidos deprehendas, et variatio reagens sine internis systematis coefficientibus immutata.

 

Munus in Hematologia Quality Control

 

Comprobatio cotidie comprobatio:Monitores constantiam computantes per canales fluidos, opticos, et impedimenta priora patienti specimen experiendi.

Analytica Error Deprehensio:Adiuvat in identitate signi calliditate, depravatione reagens, corruptioni luminis, et aperturae obstruendo.

Obsequium regulatory:Documenta qualitatem referentiae praebet ad protocolla accreditationem laboratorium sustinendum.

Sorte Transitus Validation:Mensuratio- nem baseline continuum aestimat per massam gerentem mutationes vel exercitationes systematis sustentationis.

 

Imperium gradus / Application

 

Low Level:Accuratio mensurae aestimat, fines deprehendendi inferiores, et linearitatem in leukopenicis, thrombocytopenicis, et iugis anemicae.

Normalis Level:Monitores instrumenti perficiendi intra vexillum physiologicum intervalla referentia.

Altus gradus:Fines linearitatis superiores aestimat et accurate in leukocytosi, thrombocytosi et polycythemia iugis numerat.

 

Calibrator imperium vs

 

Aspect

Hematologia Imperium

Hematologia Calibrator

Primarium munus

Aestimat ongoing analyticae praecisione

Ponit baseline mensurae scale

Supplicium Frequency

Cotidiana aut inter se operating subcinctus

Periodicum vel post- sustentationem

assignata Target

Statistical range cum superioribus et inferioribus limitibus

Certa referat valorem

Materia publica

Tri- cellularum suspensio gradu stabilita

Unius purificati- gradus referentiae

Ratio Interaction

Renuntiationes current error sine mutatione occasus

Aptat internus calibration factors

 

Compatible Hematology Analysers

 

Forma phialae apertae et clausae sampling modos trans automated tres- calculis differentialibus ex parte cellae differentialis utens modulis suggestis isotonic diluentibus.

 

Pretio assignata / Morbi

 

Parameter

Unitas

Low Level (Mean ± SD)

Normalis Level (SD Mean ±)

Princeps Level (SD Mean ±)

WBC

10^9/L

2.1 ± 0.4

7.2 ± 0.8

19.5 ± 1.8

RBC

10^12/L

2.35 ± 0.20

4.40 ± 0.30

5.50 ± 0.40

HGB

g/dL

6.2 ± 0.4

13.5 ± 0.6

17.8 ± 0.8

MCV

fl*

68.0 ± 4.0

88.0 ± 4.0

100.0 ± 5.0

PLT

10^9/L

75 ± 15

210 ± 30

480 ± 50

LYM%

%

15.0 ± 5.0

28.0 ± 6.0

50.0 ± 8.0

MED%

%

6.0 ± 3.0

8.0 ± 3.0

12.0 ± 4.0

GRAN%

%

79.0 ± 8.0

64.0 ± 7.0

38.0 ± 6.0

 

Product Specifications

 

Property

Technical Specification

Forma corporis

Liquoris suspensionis elementorum humanorum et mammalium cellularum stabilitum

Pack configurationis

Tri- Level Pack (1 Low, 1 Normal, 1 High) vel singulae sarcinae

Imple Volume

3.0 mL per phialam

Clausa- Vitalis Vitalis

CV dies a faciens diem

Aperi- Stabilitatem ampullae

XIV dies sustentari II gradus ad VIII gradus

Temperatus dolor

Repono ac sit amet in II gradus ad VIII gradus

 

Repono & Stabilitas

 

Repono Temperature:Continenter inter 2 gradum et 8 gradum. Non subjecta temperaturae torpore.

Luce Patefacio:Reponunt phialas intus originalis exterioris packaging a directo lumine nuditatis.

Mixtio Protocollum:Patitur phialam ad spatium temperaturae ambientis attingere (15 gradus ad 30 gradum) pro 15 minutis ante permixtionem. Volvuntur phialam inter palmas horizontaliter et leniter invertunt, donec resuspendatur faecis. Strenuus ne mechanica spumans ne quatiat.

 

Quality Assurance

 

Salus Testis:Unitates sanguinis donatoris negans pro HBsAg, anti-HCV tentant, et anti-HIV elementorum.

Valorem adsignatione:Valores constituuntur utentes normae referentiae methodi et calculis laboratorium calibratum.

Vestibulum Ratio:Formatae et involucri intra facilities operandi sub ISO 13485 systemata qualitatum in vitro diagnostica producta.

 

FAQ

 

Q: Quid est vita clausa- fasciae phialae?

A: Materia obtinet scopum valorem stabilitatem per 105 dies ab productione conservatam inter 2 gradum et 8 gradum.

Q: Quamdiu materia utilis manet post foramen?

A: phialam apertam stabilitatem obtinet per 14 dies vel 14 introitus sampling, cum statim recipiatur et ad 2 gradum ad 8 gradum repositorium rediit.

Q: Potestne productum congelatum ad extendere fasciae vitam?

A: Congelantia damna structurae cellulosae et lysin cellularum causat, materiam invalidam reddens pro accuratis comitibus.

Q: Quid est recta praeparatio ante aspirationem?

A: Aequilibrate phialam ad cella temperiem per 15 minuta, dein leniter inter palmas et finem inverti - supra - finem, donec cellularum sedimentum plene suspensa sine bullis aereis.

Q: Cur scopum parametri praeditum ut valoris range loco unius digiti?

A: Parametri scopum includunt limitem superiorem et inferiorem ad rationem variationum operationalium expectatarum per varios canales numerandi, systemata optica et ambitus mensurae ambientium.

Hot Tags: Hematologia analyser III-partium moderatio, Sina 3-pars hematologia analyser sistens fabricatores, praebitores

mitte Inquisitionis

whatsapp

Phone

E-mail

Quaestiones

lapides sacculi